GE Healthcare, Uni-Circuit Kit, Adult, Disposable, 1.8 m/72 in, REF M1089028, M1087972, M1087975, M1126445, M1127508, M1
FDA device recall Z-1870-2009 · posted 2009-09-09 · Terminated
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- Breathing circuit inner tube may separate from the rigid connectors under certain conditions. A separation of the inner tube from the connector may prevent adequate gas flow (O2, N2O and anesthetic agents) to the patient. A separation can also cause CO2 re-breathing. This could lead to hypoxia and/or hypercarbia and inadequate anesthesia delivery.
- Action taken
- GE Healthcare sent consignees 6/5/09 an "Urgent Medical Device Correction" letter dated June 5, 2009. The letter was addressed to Health Care Administrator/Risk Manager; Chief of Anesthesia; Director of Clinical Engineering; Risk Manager. The letter described the Safety Issue, Affected Product Details, Safety Instructions (included "Recall Confirmation" form fax to 973-790-4150), Product Correction and Contact Information.
- Root cause
- Process control
- Status
- Terminated
- Product code
- CAI
- Quantity affected
- 85097
- Distribution
- Worldwide distribution: USA, AUSTRALIA, DENMARK, FRANCE, ESTONIA, GERMANY, ITALY, JORDAN, NETHERLANDS, NORWAY, SWEDEN, and UNITED ARAB EMIRATES.
- Date initiated
- 2009-06-05
- Date posted
- 2009-09-09
- Date terminated
- 2014-01-16
- Location
- Waukesha, WI