GE Healthcare STe (16 slice) System; Discovery STe (16 Slice) Systems with technical Reference Manual numbers 2351785-xx

FDA device recall Z-0944-2011 · posted 2011-02-07 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
Although the system x-ray output is accurate for the technique factors given, and the dose display is also accurate, certain CTDI adjustment factors used for the manual calculation from the Technical Reference Manual (TRM) are incorrect. CTDI100 Aperture Adjustment Factors for 4x0.625 mode contain incorrect values. This issue may lead to inaccurate CTDI manual calculation. The results of the diffe
Action taken
An Important Electronic Product Radiation Warning letter will be provided together with the TRM addendum informing the customers of the issue and related hazards. The letter also identified the affected product and gave instructions, which include referring to the TRM addendum, along with the product correction. This correction will be executed via a Field Modification Instruction IFMI 254281 released in November, 2010 with an estimated completion date of May, 2011. Questions or concerns should be directed to the call numbers listed or a local GE Healthcare Service Representative for countries not listed.
Root cause
Radiation Control for Health and Safety Act
Status
Terminated
Product code
JAK
Quantity affected
1450 units total in US
Distribution
Worldwide Distribution
Date initiated
2010-10-27
Date posted
2011-02-07
Date terminated
2012-07-13
Location
Waukesha, WI