GE Healthcare, Revolution XR/d. Revolution XR/d Intended Use: The Revolution XR/d Digital Radiographic imaging system is

FDA device recall Z-1989-2014 · posted 2014-07-02 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
GE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
Action taken
GE Healthcare sent an Urgent Medical Device Correction letter dated June 4, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions in regard to this recall the consignees in the United States were advised to call 800-437-1171. For other countries contact your local GE Healthcare Service Representative. For questions regarding this recall call 800-437-1171.
Root cause
Equipment maintenance
Status
Terminated
Product code
KPR
Quantity affected
1053
Distribution
Worldwide Distribution - USA including PR. DC, GUAM, all states except VT. Internationally to the following countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BENIN, BOSNIA AND HERZE
Date initiated
2014-06-04
Date posted
2014-07-02
Date terminated
2015-05-06
Location
Waukesha, WI