GE Healthcare, Invasive Blood Pressure Care Cables. Product ID Number: 2016995-001, 2016995-002, 2016995-003, 2016995-00
FDA device recall Z-2015-2012 · posted 2012-07-16 · Terminated
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- GE Healthcare has become aware through complaints of a potential safety issue with its Invasive Blood Pressure Cable made for use with a Spectramed / BD invasive blood pressure transducer. Due to use of an incorrect O-ring, the mating of the cable with the transducer is faulty. This situation inhibits blood pressure measurement from being monitored correctly.
- Action taken
- GE Healthcare sent a Urgent Medical Device Correction letter dated June 12, 2012, to all affected customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Inforamtion. Advised customers to discontinue use, further distribution and to contact their customers. Please return any defective Invasive Blood Pressure Cables by isolating all affected product and completing the attached Confirmation form/fax back per its instructions. This step is required to confirm receipt of communications with all customers. If you have any questions or concerns regarding this notification, please contact Customer Service at +1-800-932-0760 (Domestic & International). Hours of Operation: 8:00 am EST to 6:00 pm EST. .
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MHX
- Quantity affected
- 279
- Distribution
- Worldwide Distribution -- USA (nationwide) including the states of AK, CA, IL, MS, NJ, PR, SD, TN, TX, and VA. and the countries of ARGENTIAN, AUSTRALIA, BELGIUM, CHINA, GERMANY, ALGERIA, ECUADOR, EGY
- Date initiated
- 2012-06-12
- Date posted
- 2012-07-16
- Date terminated
- 2013-12-05
- Location
- Waukesha, WI