GE Healthcare Innova 4100 and Innova 4100 with Bolus Chasing offered as an option. Digital Fluoroscopic Imaging System.

FDA device recall Z-0898-06 · posted 2006-05-24 · Terminated

Recall details

Recalling firm
GE Healthcare
Reason for recall
Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient.
Action taken
GEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.
Root cause
Other
Status
Terminated
Product code
MQB
Quantity affected
291 (213 domestic and 78 OUS)
Distribution
Nationwide including Puerto Rico and OUS to include: Australia, Belgium, Brazil, Bulgaria, Byelarus, Canada, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece,
Date initiated
2006-03-06
Date posted
2006-05-24
Date terminated
2008-02-24
Location
Waukesha, WI