GE Healthcare, Innova 2121IQ, Innova 3131IQ Cardiovascular X-ray imaging systems. K060259: The Innova 3131IQ (3131-IQ) a
FDA device recall Z-2376-2012 · posted 2012-09-13 · Terminated
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- GE Healthcare has recently become aware of a potential safety issue with respect to the Innova 2121IQ & Innova 3131IQ Systems. While performing a fluoroscopic acquisition with the Innova 2121IQ or Innova 3131IQ systems, there is a potential for loss of usable imaging mode which requires the system to restart in order to recover.
- Action taken
- GE Healthcare sent an Urgent Medical Device Correction letter dated August 16, 2012, to all affected consignees. The letter described the Safety Issue, safety Instructions, Affected Product Details, Product Correction, and Contact Information. Consignees were instructed that hospital staff should wait for Photo or Image to appear on the display (when store photo function is performed) before start of the next short duration fluoroscopy acquisition. For questions customers were instructed to call: United States (800) 437-1171 Japan: 0120 - 055 - 919 For other countries consignees should contact their local GE Healthcare Service Representative. For questions regarding this recall call 262-513-4122.
- Root cause
- Software design
- Status
- Terminated
- Product code
- MQB
- Quantity affected
- 240 (133 US and 107 OUS)
- Distribution
- Worldwide Distribution - USA including AL, AZ, CA, CO, CT, FL, HI, IK, IL IN, IA, KS, LA, MD, MA, MI, MN,MO, NV, NH, NH, NM, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI,.and internationally
- Date initiated
- 2012-08-16
- Date posted
- 2012-09-13
- Date terminated
- 2015-10-20
- Location
- Waukesha, WI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| GE INNOVA SOLID STATE XRAY-IMAGER FLUOROSCOPIC X-RAY SYSTEM, MODELS INNOVA 4100, 3100, 2100, 3131, 2121 (K091658) | GE Healthcare | 2009-06-29 |
| INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT (K061163) | Ge Medical Systems, LLC | 2006-06-07 |
| DIGITAL FLUOROSCOPIC IMAGING SYSTEMS - INNOVA 3131 & INNOVA 2121 (K060259) | Ge Medical Systems, LLC | 2006-02-22 |