GE Healthcare Innova 2100IQ. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human a
FDA device recall Z-1976-2015 · posted 2015-07-03 · Terminated
Recall details
- Recalling firm
- GE Healthcare
- Reason for recall
- GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).
- Action taken
- The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter on 5/29/2015- GEHC Ref#12231- dated May 29, 2015 to their Consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and contact Information. The customers were instructed that you may continue to use the system and avoid rapid direction changes with the gantry motions and prevent potential collisions with other equipment. If the failure occurs during interventional examination, please ensure that you have established procedures for handling patients in case of the loss of gantry motion as stated in your product labeling. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for this correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- IZI
- Quantity affected
- 13
- Distribution
- Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, GA, IL, IN, KS, MS, NV, OH, OK, OR, PA, TN, TX, VA, WA and WI; and countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHI
- Date initiated
- 2015-05-29
- Date posted
- 2015-07-03
- Date terminated
- 2015-10-28
- Location
- Waukesha, WI