GE Healthcare Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ, Innova 2121 IQ, Innova 3131 IQ , Innova 3100, Innova 4100.
FDA device recall Z-1318-2013 · posted 2013-05-14 · Terminated
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries have been reported due to this issue to date, there is risk of complications during the procedure as a result of imaging functionality loss.
- Action taken
- GE Healthcare did not send out customer letters. GE Healthcare completed the correction by replacing the faulty relay in all affected systems. As an effectiveness check, they have tracked all service requests for the field action to completion. These service requests recorded the repair of the affected units. Exceptions to completion of the service request were recorded as well. Records of field action completion will be retained per their processes and procedures. Should you have any questions or require additional information, please call (262) 548-2757.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- OWB
- Quantity affected
- 713
- Distribution
- Worldwide Distribution -- Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, GUAM, HI, ID, IL , IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NH, NM, NY, NC, ND, OH,
- Date initiated
- 2011-12-20
- Date posted
- 2013-05-14
- Date terminated
- 2014-03-13
- Location
- Waukesha, WI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT (K061163) | Ge Medical Systems, LLC | 2006-06-07 |
| INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ (K052412) | GE Healthcare | 2005-09-16 |
| INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ (K052157) | Ge Medical Systems, LLC | 2005-08-25 |
| INNOVA 2100 (K050489) | Ge Medical Systems, LLC | 2005-03-11 |
| INNOVA 4100, INNOVA 3100 (K042053) | Ge Medical Systems, Inc. | 2004-08-26 |