GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generati

FDA device recall Z-0934-2013 · posted 2013-03-11 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
Action taken
No consignee letter sent out. GE Healthcare's recall strategy was to replace the impacted cable on all affected Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ systems that were distributed. At this time, deployment of this recall is considered complete as of 9/6/2011.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
MQB
Quantity affected
48
Distribution
Worldwide Distribution-USA including the states of FL, IL, IN, NJ, NC, OK, TX, and WI, and the countries of UNITED KINGDOM, TUNISIA, SERBIA, RUSSIAN FEDERATION, NORWAY, NIGERIA, LITHUANIA, LEBANON, KO
Date initiated
2011-02-04
Date posted
2013-03-11
Date terminated
2014-01-27
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
GE INNOVA SOLID STATE XRAY-IMAGER FLUOROSCOPIC X-RAY SYSTEM, MODEL: 4100IQ, 3100IQ, 2100IQ (K092004)GE Healthcare2009-08-25
INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT (K061163)Ge Medical Systems, LLC2006-06-07
INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ (K052412)GE Healthcare2005-09-16
INNOVA 2100 (K050489)Ge Medical Systems, LLC2005-03-11