GE Healthcare Innova 2100 IQ Digital Fluoroscopic Imaging System (Versatile cardiac and Vascular imaging system). The In

FDA device recall Z-1518-2008 · posted 2008-09-30 · Terminated

Recall details

Recalling firm
Datex - Ohmeda, Inc
Reason for recall
User was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to incorrect overall dose management, resulting in unnecessary exposure and can be a contributing factor for skin injury. The alarm can be heard in the control room.
Action taken
GE Field Engineers are visiting all affected customers to apply updated software. For assistance contact Datax-Ohmeda, Inc., at 1-608-221-1551.
Root cause
Radiation Control for Health and Safety Act
Status
Terminated
Product code
MQB
Quantity affected
281
Distribution
within US: AL, AR, AZ, CA,CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. OUS to include: Alger
Date initiated
2006-08-17
Date posted
2008-09-30
Date terminated
2009-06-09
Location
Madison, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ (K052412)GE Healthcare2005-09-16
INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ (K052157)Ge Medical Systems, LLC2005-08-25
INNOVA 2100 (K050489)Ge Medical Systems, LLC2005-03-11