GE Healthcare, Infinia Hawkeye 4, Model Numbers: 5178728, ASM500060, H3000WC and H3000WD. The device is used to perform

FDA device recall Z-1121-2009 · posted 2009-04-10 · Terminated

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
The metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.
Action taken
Consignees were sent a GE "Product Safety Notification" letter dated April 2008. The letter was addressed to Hospital Administrators, and Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information. Contact your local sales or service representative for further information regarding this notification by calling 1-800-437-1171 (within the US or Canada) and 1-262-896-2890 (outside the US).
Root cause
Other
Status
Terminated
Product code
KPS
Quantity affected
122 US and 166 OUS
Distribution
Worldwide Distribution -- USA (DC and Puerto Rico) including states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY, NC, ND, O
Date initiated
2008-06-02
Date posted
2009-04-10
Date terminated
2012-05-24
Location
Waukesha, WI