GE Healthcare, GE Innova 2100 IQ Versatile Cardiac and Vascular Imaging System, (The system consists of an a monoplane p
FDA device recall Z-0136-2010 · posted 2009-11-09 · Terminated
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
- Action taken
- Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- IZI
- Quantity affected
- 60
- Distribution
- Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, PO
- Date initiated
- 2009-04-16
- Date posted
- 2009-11-09
- Date terminated
- 2015-12-03
- Location
- Waukesha, WI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT (K061163) | Ge Medical Systems, LLC | 2006-06-07 |
| INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ (K052412) | GE Healthcare | 2005-09-16 |
| INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ (K052157) | Ge Medical Systems, LLC | 2005-08-25 |
| INNOVA 2100 (K050489) | Ge Medical Systems, LLC | 2005-03-11 |