GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Patient Monitors, Dash Port2 Docking Station.

FDA device recall Z-0982-2014 · posted 2014-02-14 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor and Dash Port 2 Docking Station: DASH: 1. Dash stops communicating on the Unity Network resulting in Loss of Monitoring at the CARESCAPE" CIC Pro Central Station*. 2. Dash displays distorted ECG waveforms in combination monitoring mode. 3. Dash incorrectly becomes Time Master on the Unity Network. 4.
Action taken
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated November 27,2013. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information (Contact Technical Support at 1-800-558-7044 or your local Service Representative).
Root cause
Software design
Status
Terminated
Product code
MHX
Quantity affected
178,711 (92,123 US, 86,588 OUS)
Distribution
Worldwide Distribution: USA (nationwide) including DC and PR; and Internationally to: ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HE
Date initiated
2013-11-27
Date posted
2014-02-14
Date terminated
2016-08-25
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODIFICATION TO DASH 3000, 4000, 5000 MONITOR AND ACCESSORIES (K073462)Ge Medical Systems Information Technologies2008-01-11
DASH 3000 PATIENT MONITOR (K992929)General Electric Medical Systems Information Techn1999-11-26