GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Informati

FDA device recall Z-0891-2007 · posted 2007-06-06 · Terminated

Recall details

Recalling firm
General Electric Medical Systems Information Technology
Reason for recall
Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot .
Action taken
GE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07 that explained the short and long term solutions and that the company would be providing a software update that will rectify the issue specific to CIC Pro v4.1 software.
Root cause
Other
Status
Terminated
Product code
DSI
Quantity affected
5,190 Units
Distribution
Nationwide
Date initiated
2007-01-24
Date posted
2007-06-06
Date terminated
2011-12-10
Location
Milwaukee, WI