GE Healthcare, CARESCAPE Monitor B850, B650 and B450 (Bx50).

FDA device recall Z-1606-2015 · posted 2015-05-12 · Terminated

Recall details

Recalling firm
GE Healthcare
Reason for recall
The heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message and there may not be an alarm for asystole. Undetected asystole could result in irreversible changes in the patients condition and delayed or missed life sustaining patient treatment
Action taken
GE Healthcare sent an Urgent Medical Device Correction letter dated May 1, 2015, to all affected consignees. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Root cause
Software design
Status
Terminated
Product code
MHX
Quantity affected
110 (83 units US, 27 units OUS)
Distribution
Worldwide Distribution - US including CO, MI, WI and Internationally to NETHERLANDS, FINLAND, and BELGIUM.
Date initiated
2015-05-04
Date posted
2015-05-12
Date terminated
2015-09-15
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CARESCAPE MONITOR B450 (K132533)GE Healthcare Finland Oy2013-12-04
CARESCAPE MONITOR B650 (K131223)GE Healthcare Finland Oy2013-08-28
CARESCAPE MONITOR B850 (K131414)GE Healthcare Finland Oy2013-08-28
CARESCAPE MONITOR B650 (K102239)GE Healthcare Finland Oy2010-10-18
CARESCAPE MONITOR B850 (K092027)Ge Medical Systems Information Technologies2009-08-28