GE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coi

FDA device recall Z-0553-2016 · posted 2015-12-28 · Terminated

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue.
Action taken
Consignees were sent on 12/7/2015 a GE Healthcare "Urgent Medical Device Correction" GEHC Ref#60888 letter dated December 7, 2015. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. UPDATE: Consignees were sent a revised letter on May 2, 2016. Previously the letter only listed MR750W Surgical Suite Scanners. Because 3.0T 6 Channel Flex Coils are used with the MR750, MR750W, and 3.0T HDxT (including HD23), these devices have been added to the letter. There are no other changes to the letter. UPDATE 5/27/16: Consignees were sent a revised letter on May 19, 2016. The letter topic stated "De-rating of 3.0T 6 Channel Flex Coil when used on Discovery MR750 running DV24, DV25, or DV25.1 Application Software" . The letter described the Safety Issue, Safety Instruction, Affected Product details, Product Correction and Contact Information (same as previous letter).
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
MOS
Quantity affected
90 (54 US; 36 OUS)
Distribution
Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, C
Date initiated
2015-12-07
Date posted
2015-12-28
Date terminated
2018-11-08
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
1.5 GE 6-CHANNEL PHASED ARRAY FLEX COIL, 3.0 GE 6-CHANNEL PHASED ARRAY FLEX COIL (K110610)Ge Medical Systems, LLC2011-04-01