GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Anesthesia systems. The E

FDA device recall Z-0006-2008 · posted 2007-10-04 · Terminated

Recall details

Recalling firm
Datex - Ohmeda, Inc
Reason for recall
Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.
Action taken
Because this issue only occurs when the Disposable Multi Absorber is used in conjunction with EZchange Module, they have contacted all EZchange Module customers. It is critical to contact these customers specifically, rather than all customers shipped Disposable Multi Absorbers as this may cause unwarranted confusion with customers who do not have the EZchange Module. Consignees were sent an letter dated 7/19/2007 describing the product and problem. The letter lists a number of interim actions to implement with the anesthesia systems.
Root cause
Vendor change control
Status
Terminated
Product code
BSF
Quantity affected
389,070
Distribution
Worldwide, including USA, Australia, Belgium, Brazil, Canada, Chile, China, Denmark, Egypt, Estonia, Finland, France, Germany, India, Ireland, Italy, Jordan, Republic of Korea, Kuwait, Mexico, Netherl
Date initiated
2007-07-24
Date posted
2007-10-04
Date terminated
2015-10-16
Location
Madison, WI