GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Anesthesia systems. The E
FDA device recall Z-0006-2008 · posted 2007-10-04 · Terminated
Recall details
- Recalling firm
- Datex - Ohmeda, Inc
- Reason for recall
- Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.
- Action taken
- Because this issue only occurs when the Disposable Multi Absorber is used in conjunction with EZchange Module, they have contacted all EZchange Module customers. It is critical to contact these customers specifically, rather than all customers shipped Disposable Multi Absorbers as this may cause unwarranted confusion with customers who do not have the EZchange Module. Consignees were sent an letter dated 7/19/2007 describing the product and problem. The letter lists a number of interim actions to implement with the anesthesia systems.
- Root cause
- Vendor change control
- Status
- Terminated
- Product code
- BSF
- Quantity affected
- 389,070
- Distribution
- Worldwide, including USA, Australia, Belgium, Brazil, Canada, Chile, China, Denmark, Egypt, Estonia, Finland, France, Germany, India, Ireland, Italy, Jordan, Republic of Korea, Kuwait, Mexico, Netherl
- Date initiated
- 2007-07-24
- Date posted
- 2007-10-04
- Date terminated
- 2015-10-16
- Location
- Madison, WI