GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data within
FDA device recall Z-1333-2022 · posted 2022-07-05 · Open, Classified
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)
- Action taken
- GE notified their consignees by issuing an URGENT MEDICAL DEVICE CORRECTION notice on 08 June 2022. The letter explains the two safety issues and provides the following actions be taken: Issue #1: "You can continue to use your device in accordance with the User Manual and the actions below: Do not utilize any measurements when the 'Lossy' overlay is displayed in the lower left of the viewport and, when you hover over the word 'Lossy', it displays the tooltip box with the phrase 'Image scaled down from its original resolution'." Issue #2: "You can continue to use your device in accordance with the User Manuals and the actions below: Do not perform any measurement on non-squared pixel images. Always confirm critical diagnostic measurements by other means (i.e using a fiducial marker of known size)." GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. After the software has been corrected, be sure to destroy the installation media for affected software at your site. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- LLZ
- Quantity affected
- 1,934 units
- Distribution
- Worldwide distribution.
- Date initiated
- 2022-06-08
- Date posted
- 2022-07-05
- Location
- Waukesha, WI