GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems (an accessory

FDA device recall Z-0001-2007 · posted 2006-10-03 · Terminated

Recall details

Recalling firm
GE OEC Medical Systems, Inc
Reason for recall
Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.
Action taken
Consignees were notified by letter on 05/31/2006
Root cause
Process control
Status
Terminated
Product code
JAA
Quantity affected
143 units
Distribution
Nationwide, including one VA facility in PA. No military distribution. Foreign distribution to Canada, China, Italy, Japan, Spain.
Date initiated
2006-05-31
Date posted
2006-10-03
Date terminated
2011-07-27
Location
Salt Lake City, UT