Gas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum pati
FDA device recall Z-1896-2016 · posted 2016-06-13 · Terminated
Recall details
- Recalling firm
- Carefusion 211 Inc dba Carefusion
- Reason for recall
- CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Action taken
- CareFusion sent a Field Safety Notice dated May 17, 2016, to all affected customers. The notice informed customers that CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM). This malfunction creates a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation to the patient. By design an audible clinical alarm is activated. Customers were informed of the problems identified and the actions to be taken. Customers were not required to return the affected devices or suspend use, customers would receive an copy of the identified affected serial numbers, as well as a response card, customers were requested to promptly return the response card to expedite the correction process and acknowledge receipt of the notification, Customers would be contacted by CareFusion to arrange onsite remediation of the affected devices, in the interim if any AVEA ventilator unit in their facility exhibits a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation, immediately remove the ventilator from service, provide alternate ventilation and contact CareFusion Technical Support per the contact information listed below to report the issue. Customers with questions should call 888-562-6016, for technical support 800-231-2466 or support.vent.us@carefusion.com.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- CBK
- Quantity affected
- 922 units total (501 units in US)
- Distribution
- Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Sing
- Date initiated
- 2016-05-17
- Date posted
- 2016-06-13
- Date terminated
- 2017-09-28
- Location
- Yorba Linda, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| AVEA VENTILATOR (K103211) | Care Fusion | 2011-05-12 |
| MODIFICATION TO AVEA VENTILATOR (K073069) | Viasys Respiratory Care, Inc. | 2008-01-25 |
| AVEA (K013642) | Bird Products Corp. | 2002-05-22 |