GammaWin software, part number GM11019110, versions up to and including 1.62, for the GammaMedplus radionuclide applicat

FDA device recall Z-1268-06 · posted 2006-07-27 · Terminated

Recall details

Recalling firm
Varian Medical Systems
Reason for recall
Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment data is entered manually.
Action taken
The recalling firm notified end users by letter on 05/01/06 and advised users of possible communication error between afterloader devices and software. The notification also advised that the control program for the devices has the potential to lead to a misadministration of treatment in cases where data is manually entered by the user. The firm additionally advised that user may continue to use the afterloader devices if following the listed cautionary statements. A software upgrade is planned to be implemented at the user site''s next regularly scheduled radiation source exchange. A response form is enclosed with the notification and users are requested to return the customer receipt verification via fax.
Root cause
Other
Status
Terminated
Product code
JAQ
Quantity affected
254 total units
Distribution
Nationwide and worldwide.
Date initiated
2006-05-01
Date posted
2006-07-27
Date terminated
2007-05-18
Location
Charlottesville, VA