Atlas FDA

GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 200 mm. Remote controlled radionuclide applicator system.

FDA device recall Z-0327-2012 · posted 2011-12-01 · Terminated

Recall details

Recalling firm
Varian Medical Systems, Inc.
Reason for recall
Plastic needle tip could possibly crack or separate.
Action taken
Varian Medical Systems sent a "CUSTOMER TECHNICAL BULLETIN" letter dated September 30, 2008 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed users to inspect needle tips before and after every clinical application to prevent needle tip separation or cracking.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JAQ
Quantity affected
1716 parts
Distribution
Worldwide Distribution - USA (nationwide) including the states of Arizonia, California, Florida, Illinois, Indiana, Michigan, North Carolina, and Vermont, and the countries of Argentina, Austria, Belg
Date initiated
2008-09-30
Date posted
2011-12-01
Date terminated
2011-12-07
Location
Charlottesville, VA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
GAMMAMED PLUS HIGH DOSE RATE REMOTE AFTERLOADING SYSTEM (K983436)Gammamed USA1998-10-29