G7 Acetabular Screw 6.5mmx35mm Part Number: 010001000
FDA device recall Z-0493-2014 · posted 2013-12-13 · Terminated
Recall details
- Recalling firm
- Biomet, Inc.
- Reason for recall
- Investigation identified that the outer label, inner label, and patient label all have the incorrect diameter size of 10.5mm. The screw is actually a 6.5mm diameter and correctly corresponds to the items number on the label.
- Action taken
- Beginning November 25 Urgent recall electronic notices were sent by Biomet UK to all distributors with hospitals outside the USA notified by ground mail November 27,2013. The notification identified the product and reason for the recall, asking for immediate location and discontinued use of the product and return to Biomet UK, Ltd. Questions can be addressed to Biomet USA at 01 574-372-1570 M-F 8 am to 5 pm.
- Root cause
- Labeling Change Control
- Status
- Terminated
- Product code
- JWH
- Quantity affected
- 532 in USA all under Biomet USA control
- Distribution
- Worldwide Distribution - USA Nationwide and in the countries of Europe, Italy, UK, Australia, Japan. all product in USA is under Biomet control.
- Date initiated
- 0013-11-26
- Date posted
- 2013-12-13
- Date terminated
- 2014-09-29
- Location
- Warsaw, IN