Atlas FDA

FreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Gluc

FDA device recall Z-0485-2014 · posted 2013-12-17 · Terminated

Recall details

Recalling firm
Abbott Diabetes Care, Inc.
Reason for recall
Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.
Action taken
Abbott Diabetes Care sent an Urgent Product Recall letter in November 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who were affected were instructed to take the following steps: Call Abbott Diabetes Care Customers Service immediately at 1-888-736-9869 for replacement of the affected product. While waiting for replacement strips to arrive, use an alternate method to measure their blood glucose. If only test strips available to customers were from affected lots, customers should not stop testing their blood glucose. Customers were instructed to dispose of the test strips in accordance with their local and state disposal regulations. Customers with questions were instructed to call Abbott Diabetes Care Customers Service at 1-888-736-9869. For questions regarding this recall call 510-239-2775. Abbott issued a press release at the request of CDRH on november 27, 2013.
Root cause
Other
Status
Terminated
Product code
NBW
Quantity affected
141,219 cartons for human use - all lot numbers
Distribution
Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Australia, Austria, Belgium, Denmark, Egypt, Finland, France, Germany, Jamaica, Ireland, Israel, United Kingdom, Luxemb
Date initiated
2013-11-18
Date posted
2013-12-17
Date terminated
2015-08-03
Location
Alameda, CA