FDA device recall Z-0703-2018 · posted 2018-02-23 · Terminated

Recall details

Recalling firm
Oakworks Inc
Reason for recall
The foot control used on the OAKWORKS Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.
Action taken
OAKWORKS, Inc. issued a MEDICAL DEVICE ADVISORY NOTICE dated 8/18/2017 to customers providing more information in regards to the Medical Device Advisory Notice sent on May 12, 2017. The advisory notice was initiated in response to reported issues with the foot control. OakWorks, Inc. customer service will place the order for the appropriate Foot Control Kit listed below based on what table is impacted and the date of manufacture. PN 2015-204 - Foot Control Kit - Standard Riser 2015-209 - Foot Control Kit - Thinner Riser 2015-210 - Foot Control Kit - Holder Only 2015-211 - Foot Control Kit - Riser Removal Customers with questions may contact 1-800-558-8850 or 1-717-235-6807.
Root cause
Device Design
Status
Terminated
Product code
IXR
Quantity affected
1,104
Distribution
USA (nationwide) Distribution
Date initiated
2017-07-31
Date posted
2018-02-23
Date terminated
2019-03-29
Location
New Freedom, PA