Foam electrodes with Silver-Carbon construction: P/N 202-9271, Cat No. 400-898, 2" x 2" Electrode, Foam, AG, A

FDA device recall Z-0818-2013 · posted 2013-02-14 · Terminated

Recall details

Recalling firm
NAImco Inc dba Richmar Inc
Reason for recall
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
Action taken
Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.
Root cause
Unknown/Undetermined by firm
Status
Terminated
Product code
GXY
Quantity affected
56980 electrodes
Distribution
Distributed in the states of GA and TN.
Date initiated
2009-11-30
Date posted
2013-02-14
Date terminated
2013-09-03
Location
Chattanooga, TN