FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
FDA device recall Z-0628-2016 · posted 2016-01-11 · Terminated
Recall details
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Reason for recall
- Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.
- Action taken
- BD Biosciences sent an Urgent Product Recall letter dated December 2015, to all affected customers to inform them that Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to return the Tracking/Verification form to BD Biosciences. Customers are instructed to contact BD Customer Support at (877) 232-8995 (prompt 3, 3) in the US. Customers outside of the US are instructed to contact their local BD Biosciences or distributor.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- MVU
- Quantity affected
- 1,388 units
- Distribution
- Nationwide Distribution
- Date initiated
- 2015-12-14
- Date posted
- 2016-01-11
- Date terminated
- 2016-10-31
- Location
- San Jose, CA