FDA device recall Z-0108-2026 · posted 2025-10-10 · Open, Classified

Recall details

Recalling firm
MICROVENTION INC.
Reason for recall
Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.
Action taken
On 9/5/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop using and quarantine affected devices. 2) Provide the notice to medical facilities or users to whom you have distributed recalled devices. 3) Return effected devices. 4) Complete and return the customer acknowledgement form via email to Terumo8943@sedgwick.com Direct questions to the firm at terrence.callahan@microvention.com On 10/29/2025, additional recall notices were distributed identifying additional distributed devices.
Root cause
Device Design
Status
Open, Classified
Product code
OUT
Quantity affected
467
Distribution
Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA,
Date initiated
2025-09-05
Date posted
2025-10-10
Location
Aliso Viejo, CA