FDA device recall Z-0042-2017 · posted 2016-10-05 · Terminated

Recall details

Recalling firm
Trumpf Medical Systems, Inc.
Reason for recall
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
Action taken
Trumpf Medical sent an Urgent Field Safety Notice dated July 8, 2015, to all affected consignees. The letter indicated that the tables should not be used until an authorized service engineer has exchanged the spindles of the affected operating tables. The letter included an acknowledgement form which is to be returned. The letter also requested a sub-recall if the product had been further distributed. For questions regarding this recall call 843-329-0543.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JEA
Quantity affected
84 units
Distribution
Worldwide Distribution - US including CA, CT, GA, IL, NC, OR, SC, VA, WI, WY, and Internationally to Australia, Azerbaijan, Brazil, Finland, France, Germany, Hungary, Italy, Luxembourg, Norway, Poland
Date initiated
2016-07-10
Date posted
2016-10-05
Date terminated
2017-03-02
Location
Charleston, SC