FDA device recall Z-0754-05 · posted 2005-04-27 · Terminated

Recall details

Recalling firm
Ethicon Endo-Surgery
Reason for recall
Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.
Action taken
The recalling firm sent letters to hospitals and distributors, dated 12/14/04.
Root cause
Other
Status
Terminated
Product code
MDM
Quantity affected
96 devices.