FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Femoropopliteal SES clinical pro
FDA device recall Z-1510-2013 · posted 2013-06-06 · Terminated
Recall details
- Recalling firm
- Cordis Corporation
- Reason for recall
- Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
- Action taken
- Cordis sent an Urgent Medical Device Correction and Potential Removal letter dated April 24, 2013 e-mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected products to prevent further use of the affected product, until inspected. Use only product identified with a blue dot next to the lot number on the carton face label, carton end label, and outer pouch label, or product with upgraded sterile barrier system (not yet available, as of the date of this letter), review, complete, sign and return the enclosed Acknowledgement Form and maintain a copy of this notice with the affected product. For questions related to the issue, please contact our FSS/Cordis representative, or please contact Cordis Direct at (800) 781-0282, Monday through Friday from 9:00 AM to 5:00 PM EDT.
- Root cause
- Package design/selection
- Status
- Terminated
- Product code
- NIP
- Quantity affected
- 250 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Nether
- Date initiated
- 2013-04-23
- Date posted
- 2013-06-06
- Date terminated
- 2014-06-27
- Location
- Miami Lakes, FL