Flexor Tuohy-Borst Side-Arm Introducer (Shuttle Select), RPN KSAW-5.0-38-90-RB-SHTL-FLEX-HC, Order Number G31133 Flexor

FDA device recall Z-0610-2021 · posted 2020-12-23 · Terminated

Recall details

Recalling firm
Cook Inc.
Reason for recall
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
Action taken
The firm, Cook Medical, sent "URGENT: MEDICAL DEVICE RECALL" letters dated 11/24/20 to customers. The letter described the product, problem and actions to be taken. The customers were instructed to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This medical device recall letter must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing devices potentially impacted from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue. Other Information This action is being taken with the knowledge of the Food and Drug Administration. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: " Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to
Root cause
Process control
Status
Terminated
Product code
DYB
Quantity affected
902
Distribution
US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS,
Date initiated
2020-11-24
Date posted
2020-12-23
Date terminated
2022-09-07
Location
Bloomington, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Flexor Tuohy-Borst Side-Arm Introducer, Ansel Modification; Flexor Check-Flo Introducer, Ansel Modification; Flexor Check-Flo Introducers, Ansel Modification with High-Flex Dilator and Hydrophilic Coating; Flexor Check-Flo Introducer Balkin Up and Over Contralateral Design; Flexor Up and Over Balkin Contralateral Introducer; Flexor Check-Flo Introducer, Raabe Modification; Flexor Check-Flo Performer Introducer (K142829)Cook Incorporated2015-07-28