Flexiflo Quantum Enteral Pump Model # 50596, 50598 and 59322. The Flexiflo Quantum enteral pump is a medical device that
FDA device recall Z-2252-2009 · posted 2009-09-21 · Terminated
Recall details
- Recalling firm
- Abbott Nutrition, Abbott Laboratories
- Reason for recall
- The firm initiated this voluntary recall because some of its Flexiflo Quantum Enteral Pumps, have a 110-volt AC power cord plug that may crack or fail. This potential power cord plug failure may cause electrical shock and/or a delay in feeding that could represent a potential health and safety risk.
- Action taken
- An "Urgent Device Recall" letter dated September 4, 2009 was issued via FedEx overnight delivery to all direct consignees. The letter described the affected product, issue and immediate actions for customers. Direct consignees were further instructed to immediately contact affected customers and notify them about the Urgent Device Recall. Abbott Nutrition (AN) Device Call Center (DCC) will telephone those customers who do not responded promptly to the "Urgent Device Recall" letter with the completed AN Account Summary of the Flexiflo Quantum Enteral Pump Shipments. AN Device Call Center will make arrangements for the service or replacement of the recalled pumps. Please contact the AN Device Call Center at 1-877-457-0249, Monday through Friday between the hours of 8:30 am and 5:00 pm Eastern Standard Time (EST). An AN representative will work with you to arrange service or replacement of the recalled enteral pumps.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LZH
- Quantity affected
- 5,173 pumps
- Distribution
- Nationwide Distribution (states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, T
- Date initiated
- 2009-09-04
- Date posted
- 2009-09-21
- Date terminated
- 2012-09-19
- Location
- Columbus, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| FLEXIFLO QUANTUM ENTERAL NUTRITION PUMP (K913958) | Abbott Laboratories | 1992-02-07 |