Atlas FDA

Fisher & Paykel IW930 CosyCot Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module

FDA device recall Z-0485-2007 · posted 2007-01-11 · Terminated

Recall details

Recalling firm
Fisher & Paykel Healthcare Ltd
Reason for recall
This firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height. A bent pivot plate generally results in the Warmer being unable to elevate properly. Other component failures may result.
Action taken
Fisher & Paykel Healthcare commenced sending out Product Correction Notifications letter via U.S. Postal Service certified mail October 9, 2006. This Product Correction Notice explained the nature of the problem and include a reminder of the maximum recommended weight of accessories. Two FAX Response Forms were provided, Part A for initial receipt of the Product Correction Notice and Part B for confirmation of the completed upgrade.
Root cause
Other
Status
Terminated
Product code
FMT
Quantity affected
273 units in US & 9 in Canada
Distribution
Nationwide and Canada
Date initiated
2006-10-09
Date posted
2007-01-11
Date terminated
2012-04-26
Location
Auckland, New Zealand

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INFANT RADIANT WARMER - MODELS IW910 AND IW920 (K970432)Fisher & Paykel Electronics , Ltd.1997-05-07