Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a spinal fixation construct. Rx Onl

FDA device recall Z-1881-2016 · posted 2016-06-01 · Terminated

Recall details

Recalling firm
Orthofix, Inc
Reason for recall
There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle.
Action taken
Affected distributors will be provided a written notification of field correction via USPS Certified Mail or FedEx on 4/27/16. Distributors will be additionally instructed to provide field correction notifcations to hospitals, surgeons, or other distributors that they may have further ditributed the product to.
Root cause
Device Design
Status
Terminated
Product code
HXX
Quantity affected
36 units
Distribution
KS, NV, TX, CA, ID, FL, MA, MI, NJ, AZ, NH, SC, UT, WA
Date initiated
2016-04-27
Date posted
2016-06-01
Date terminated
2016-11-23