''Fine Tuning Your Pump'' booklet provided with Model MMT-515 AND MMT-715 INSULIN INFUSION PUMPS

FDA device recall Z-0162-06 · posted 2005-11-18 · Terminated

Recall details

Recalling firm
Medtronic MiniMed
Reason for recall
The text description of how to perform insulin sensitivity testing and associated worksheet included in the booklet that is shipped with the Paradigm 515 and Paradigm 715 insulin pumps contained a basic flaw in that the blood glucose after the correction should be compared to the target glucose and it says compare to the beginning blood glucose in the procedure.
Action taken
All end users who recieved the model MMT-515 or MMT-715 pump that included the 'Fine Yuning Your Pump' workbook will be notified through a letter.
Root cause
Other
Status
Terminated
Product code
LZG
Quantity affected
Approximately 14,560
Distribution
Nationwide, Bahamas, Bermuda, Canada, Mexico, and Puerto Rico.
Date initiated
2005-04-27
Date posted
2005-11-18
Date terminated
2007-03-20
Location
Northridge, CA