FilmArray Pneumonia Panel (Pneumo), REF: FIT-ASY-0144, CED 2797, IVD, Rx Only (30 test kit)
FDA device recall Z-0994-2024 · posted 2024-02-06 · Open, Classified
Recall details
- Recalling firm
- BioFire Diagnostics, LLC
- Reason for recall
- Their is a potential for false positive results when using Pneumonia panel.
- Action taken
- On December 14, 2023, Biomerieux, issued a "Urgent Medical Recall" to affected consignees via: FedEx . Biomerieux asked consignees to take the following actions: 1. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. 2. Use another method to confirm positive results for seasonal Coronavirus from the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus prior to reporting the result (e.g., consider results from testing a nasopharyngeal swab with a different BIOFIRE respiratory panel such as the BIOFIRE Respiratory Panel 2.1). Alternatively, the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus seasonal Coronavirus positive result should not be reported. 3. Complete the Acknowledgement Form in Attachment A (on the following page) and return it to bioMrieux to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this urgent product correction notice does not impact your facility.
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- QBH
- Quantity affected
- 274,290 units
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR
- Date initiated
- 2023-12-14
- Date posted
- 2024-02-06
- Location
- Salt Lake City, UT
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| FilmArray Pneumonia Panel (K180966) | Biofire Diagnostics, LLC | 2018-11-09 |