Atlas FDA

FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or D

FDA device recall Z-0895-05 · posted 2005-06-16 · Terminated

Recall details

Recalling firm
FHC, Inc.
Reason for recall
Potential for non-sterile pin to contaminate sterile field
Action taken
FHC notified customers by letter dated 3/14/05 and 3/25/05. The firm has issued a temporary set of instructions which include covering the non-sterile engagement pin with sterile tape and an amended Instruction For Use page to all customers. A redesigned pulley cover which will be sent to all customer when it becomes available in April 2005.
Root cause
Other
Status
Terminated
Product code
HAW
Quantity affected
169 units
Distribution
Natiowide
Date initiated
2005-03-14
Date posted
2005-06-16
Date terminated
2008-07-08
Location
Bowdoinham, ME