Femoral Impactor ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee r
FDA device recall Z-0905-2015 · posted 2015-01-05 · Terminated
Recall details
- Recalling firm
- DePuy Orthopaedics, Inc.
- Reason for recall
- ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
- Action taken
- DePuy Synthes sent an Urgent Information letter dated November 10, 2014, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to complete the Hospital/User Facility Acknowledgement Car and return to their Deputy Orthopaedics Sales Consultant or fax to 574-371-4939. Customer swere also asked to retain a copy of the complete Acknnowlegement Card in their files along with the notice. Customers with questions were instructed to call 574-371-4917 (M-F; 8 a.m. 5 p.m. EDT). For questions regarding this recall cal 574-371-4917.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LXH
- Quantity affected
- 4661
- Distribution
- Worldwide Distribution: US (nationwide) including P.R. except AK and RI; and Internationally to: Chile, Canada, Australia, Austria, Belgium, China, France, Ireland, India, Israel, Italy, Japan, Malays
- Date initiated
- 2014-11-10
- Date posted
- 2015-01-05
- Date terminated
- 2015-12-08
- Location
- Warsaw, IN