Femoral Artery Pressure Monitoring Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen un

FDA device recall Z-0801-2016 · posted 2016-02-24 · Terminated

Recall details

Recalling firm
Cook Inc.
Reason for recall
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.
Action taken
On 1/6/2016, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Management/Recall Administration via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
Root cause
Process control
Status
Terminated
Product code
DQY
Quantity affected
1515
Distribution
Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada
Date initiated
2016-01-06
Date posted
2016-02-24
Date terminated
2017-08-23
Location
Bloomington, IN