FC 500 MPL Flow Cytometry System with MXP Software Version 2.0 & 2.1, Part Number 626554 FC 500 MPL 733313 MPL Hardware
FDA device recall Z-0317-2007 · posted 2006-12-27 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc
- Reason for recall
- The ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. Issue limited only to protocols utilizing the Ratio parameter such as LeukoSure and DNA Cell Cycle.
- Action taken
- A Product Corrective Action (PCA) letter was mailed on Oct 17, 2006 to FC500 MPL and MXP software accounts. They are informed that if the operator opens the Parameter Selection Dialogue box after parameters have been selected for the RATIO Parameter, the selected parameters could revert back to the default choices. If these same parameters were used in the creation of the RATIO Parameter in any other saved protocol, the protocols will also be changed once they are opened after the error occurs. An immediate user preventive action to remedy this issue each time acquisition of samples that require a protocol utilizing the ratio parameter is desired is given within the letter. Consignees are asked to complete and return the enclosed Response Form within 10 days.
- Root cause
- Other
- Status
- Terminated
- Product code
- GKZ
- Quantity affected
- 37 units in the US; 4 units in Canada
- Distribution
- Nationwide and Canada
- Date initiated
- 2006-10-17
- Date posted
- 2006-12-27
- Date terminated
- 2012-02-29
- Location
- Brea, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.