FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401
FDA device recall Z-2278-2025 · posted 2025-08-08 · Open, Classified
Recall details
- Recalling firm
- Boston Scientific Corporation
- Reason for recall
- The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.
- Action taken
- An Urgent Medical Device Removal - Immediate Action Required notification letter dated June 2025 was sent to customers. Actions 1. Do NOT use affected product (Attachment 1), remove affected devices from your facility s inventory, segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page five. 5. Return affected product. Email: BSCFieldActionCenter@bsci.com Reach out to your local Boston Scientific representative with any questions.
- Root cause
- Nonconforming Material/Component
- Status
- Open, Classified
- Product code
- QZI
- Quantity affected
- 445 units
- Distribution
- US Nationwide.
- Date initiated
- 2025-06-25
- Date posted
- 2025-08-08
- Location
- Saint Paul, MN
Monitor recalls programmatically