Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
FDA device recall Z-0239-2023 · posted 2022-11-15 · Open, Classified
Recall details
- Recalling firm
- Getinge Usa Sales Inc
- Reason for recall
- The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.
- Action taken
- On October 14, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were provided with extra precautions to take in order to continue use of the device. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 943-8872 (select option 2), Monday through Friday, 8AM - 6PM ET.
- Root cause
- Equipment maintenance
- Status
- Open, Classified
- Product code
- FQO
- Quantity affected
- 5 US; 1008 OUS
- Distribution
- Worldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN. The countries of Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Bulgaria (BG), Canada (CA), China (CN), Croatia (H
- Date initiated
- 2022-10-14
- Date posted
- 2022-11-15
- Location
- Wayne, NJ