Extended Tip Applicator, 8CM, Box of 5.

FDA device recall Z-1958-2025 · posted 2025-06-20 · Open, Classified

Recall details

Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Reason for recall
Devices contain potentially out-of-specification levels of endotoxins and lack of sterility assurance.
Action taken
An URGENT: VOLUNTARY MEDICAL DEVICE RECALL notification dated 5/22/25 was mailed to consignees for this product removal. Consignees are instructed to inspect their inventory for affected devices and quarantine any identified. The provided Acknowledgement form is to be completed and returned to Integra by email at FCA2@integralife.com or by FAX at 1-609-750-4220. The provided notification is to be forwarded to users of affected devices and a copy retained for consignee records. A Return Material Authorization number will be generated upon receipt of the Acknowledgment Form for credit. Sales Representatives and Distributors are instructed to inspect their inventory for affected devices and quarantine any identified. The provided Acknowledgement form is to be completed and returned to Integra by email at FCA2@integralife.com or by FAX at 1-609-750-4220. Distributors who further distributed affected devices are to forward the provided notification to their customers and collect Acknowledgement forms on their behalf to submit to Integra. Sales Representatives will be contacted by Post Market Quality to coordinate return of devices. Consignees with any questions are to contact Customer Service by phone at 1-800-654-2873 or by email at custsvcnj@integralife.com. Additional communications were made to customers on 7/10/25 and 7/18/25 informing them that expired devices are also subject to this recall and should be disposed of.
Root cause
Process change control
Status
Open, Classified
Product code
FMF
Quantity affected
6,216 units
Distribution
Domestic: Nationwide Distribution; Foreign: Canada, Japan, Australia, New Zealand, Taiwan (Province of China), Mexico, Colombia, Dominican Republic (the), Singapore, Mongolia, Hong Kong, Chile, Argent
Date initiated
2025-05-22
Date posted
2025-06-20
Location
Princeton, NJ