Extended Brilliance Workstation (EBW) a component of the Computed Tomography X-Ray System Model Numbers EBW (CT) 728260
FDA device recall Z-1819-2011 · posted 2011-03-24 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- During an FDA inspection on 10/1/2010 the firm was notified by FDA that their field action they initiated in Oct 2009, constituted a recall. The firm determined that the functional analysis feature in the Comprehensive Cardiac Analysis (CCA) was displaying incorrect values on the polar fields. The calculated results for the myocardial wall thickening was incorrect in that the numbers displayed f
- Action taken
- Philips sent an URGENT - Field Safety Notice letter dated September 29, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to be aware that the numbers displayed for "Wall Thickening" actually represent a reduction in wall thickness from ES to ED. Philips will be providing a sotware update to resolve the issue. For further information or support customers were instructed to contact their local Philips representative: Customer Care Service Center in the U.S. at 1-800-722-9377, option 5: Diagnostic Imaging option 1: CT or their local Philips representative.
- Root cause
- Other
- Status
- Terminated
- Product code
- KPS
- Quantity affected
- 2,068 units
- Distribution
- Worldwide Distribution - USA including the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA,
- Date initiated
- 2009-10-03
- Date posted
- 2011-03-24
- Date terminated
- 2012-09-21
- Location
- Cleveland, OH