ExacTrac 6.0.x Patient Positioning System, Radiation therapy. Intended to be used to place patients at an accurately def

FDA device recall Z-2440-2015 · posted 2015-08-21 · Terminated

Recall details

Recalling firm
Brainlab AG
Reason for recall
ExacTrac 6.0 Patient Positioning System: Display of potentially incorrect Digitally Reconstructed Radiograph (DRR) for x-ray correction and verification.
Action taken
Brainlab intends to distribute the Field Safety Notice / Product Notification letter CAPA-20150713-001446, dated July 20, 2015, to the customers. The letter provides instructions for the user on how to address this issue, and to avoid a patient and treatment target position potentially different than intended due to this issue. The Field Safety Notice / Product Notification letter states that as generally required, always verify that the correct DRRs for the current patient are displayed in ExacTrac and an accurate fusion to the X-Ray images is possible for patient positioning. In case incorrect DRRs are displayed, exit and re-start the ExacTrac software. To minimize the probability of this issue occurring for the affected ExacTrac versions: -Reduce the amount of contours exported to ExacTrac as far as possible. -Avoid exporting large contours, such as couch top models, to ExacTrac. -After every (re-)start of the ExacTrac application on any workstation, always first load a phantom plan not containing contours. The Field Safety Notice / Product Notification letter states that as a general reminder the user should always follow the instructions and warnings as described in the user guide. In the context of this notification please specifically consider the safety notes relevant for X-ray correction and verification: -The image overlay functions must be used to verify fusion accuracy in both image views, especially if the images contain a series of similar structures such as vertebrae. -Do not perform patient treatment unless accurate image fusion is possible. -Large correction shifts (e.g. greater than 10 mm) may indicate incorrect automatic fusion. In such cases, verify the correction shift a second time. Brainlab Corrective Action: 1. Existing potentially affected ExacTrac 6.0.x customers receive the product notification information. 2. Brainlab will provide a software revision (ExacTrac v. 6.0.6) with this issue solved to affe
Root cause
Software in the Use Environment
Status
Terminated
Product code
IYE
Quantity affected
361 systems
Distribution
Worldwide Distribution - USA including the states of AL, AZ, AK, CA CO, CT, FL, GA, ID, IL IN, KY, LA, MD, MA, MI, MN, NE, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, and WI; and, the countrie
Date initiated
2015-08-05
Date posted
2015-08-21
Date terminated
2017-01-05
Location
Feldkirchen, Germany