Atlas FDA

ExactaMix EVA Container, 250mL, REF: 737, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to

FDA device recall Z-1963-2011 · posted 2011-04-12 · Terminated

Recall details

Recalling firm
Baxa Corporation
Reason for recall
TPN bags may leak during filling operation, due to bag manufacturing error.
Action taken
The firm, BAXA, sent a "ExactaMix Bag Product Notice" dated March 8, 2011 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: If they do not intend to use the manual addition port, verify that the cap is firmly seated onto the port by gently tugging on the cap. If they do intend to use the manual addition port, they should snap off the additive port cap prior to compounding the bag. If they encounter a faulty additive port, contact Baxa Technical Support to receive credit and/or replacement on defective product only. To return affected bags, follow the instructions provided in the notice if the bags were purchased directly from Baxa or purchased from wholesaler, distributor or other 3rd party vendor. Additional copies of the return form can be retrieved by going to www.baxa.com/ebug OR by calling Baxa Technical Support at 800-678-2292. If you have any questions or concerns, contact the Sr. Vice President, Quality and Regulatory at (303) 400-2838 or Baxa Technical Support at 800-678-2292.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
LHI
Quantity affected
733,400 bags
Distribution
Worldwide distribution: USA (nationwide) and countries including: Argentina, Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, Hong Kong, Ireland, Japan, Malaysia, Mexico, New Zealand, Panam
Date initiated
2011-03-08
Date posted
2011-04-12
Date terminated
2012-03-23
Location
Englewood, CO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP (K900585)Baxa Corp., Sub. of Cook Group, Inc.1990-04-18