Atlas FDA

Exacta-Mix 2400, Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on fin

FDA device recall Z-1370-2008 · posted 2008-08-19 · Terminated

Recall details

Recalling firm
Baxa Corporation
Reason for recall
Flush solution from TPN compounder may be added to patient's TPN bag.
Action taken
Consignees were notified by letter (Exacta-Mix Compounder Software Safety Alert) on 02/11/2008 advising them of the necessity to install a waste bag prior to a machine flush.
Root cause
Software design
Status
Terminated
Product code
NEP
Quantity affected
1000 units
Distribution
Nationwide, including military facilities in CA, FL, HI, MS, NC, TX, VA and VA facilities in NY, OR, SC, UT. Foreign distribution to Australia, Canada, China and New Zealand.
Date initiated
2008-02-11
Date posted
2008-08-19
Date terminated
2012-03-16
Location
Englewood, CO