EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V
FDA device recall Z-2467-2023 · posted 2023-08-28 · Open, Classified
Recall details
- Recalling firm
- Olympus Corporation of the Americas
- Reason for recall
- Recent reports of patient infection.
- Action taken
- Consignees were sent an Urgent Medical Corrective Action notification dated 7/27/2023. This action is being initiated due to the firm becoming aware of and FDA inquiring about recent reports of infections and positive cultures. Olympus is reminding users to closely follow reprocessing instructions, including periodic inspections by Olympus Service, and informing users about an updated manual intended to improve comprehension of existing reprocessing steps. Additional enhancements to the manual include the use of reprocessing timelines and simplified text/visualization techniques. The manual is available on the customer web portal and paper copies, along with wall charts, are available to be mailed to your facility. The notification also emphasises the required inspection of the forceps elevator mechanism, to be performed once a year or every 100 reprocessing cycles, and encourages consignees to consider incorporating voluntary routine microbiologic surveillance of endoscops as part of their quality improvement/assurance efforts. Consignees are asked to complete and return the provided response form to Sedgwick via email at ollympus7521@sedgwick.com or by fax to 888-208-4588. Questions can be directed by phone to 866-912-9552. Customers can also contact Ashley Mitch at Olympus by phone at 484-553-1029 or by email at Ashley.Mitch@olympus.com with any additional concerns.
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- FDT
- Quantity affected
- 6,426 units
- Distribution
- US Nationwide distribution.
- Date initiated
- 2023-07-27
- Date posted
- 2023-08-28
- Location
- Center Valley, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| Evis Exera III Duodenovideoscope Olympus TJF-Q190V (K220587) | Olympus Medical Systems Corp. | 2022-05-11 |
| Evis Exera III Duodenovideoscope Olympus TJF-Q190V (K202661) | Olympus Medical Systems Corp. | 2020-12-08 |
| Evis Exera III Duodenovideoscope Olympus TJF-Q190V (K193182) | Olympus Medical Systems Corp. | 2020-01-17 |