Atlas FDA

ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycopla

FDA device recall Z-0633-05 · posted 2005-03-18 · Terminated

Recall details

Recalling firm
Diasorin Inc.
Reason for recall
2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50. The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail.
Action taken
Customers were sent notification letters. Recommended to detroy the remaining kits and requested a form be returned which indicates the number of kits destroyed so they can arrange replacement.
Root cause
Other
Status
Terminated
Product code
LAZ
Distribution
MA, OH & IL
Date initiated
2004-10-20
Date posted
2005-03-18
Date terminated
2006-07-22
Location
Stillwater, MN

Monitor recalls programmatically